Open, Classified

INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 32mm, REF 3604532; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.

FDA Recall Z-1410-2023 · Posted 2023-04-19

Recall StatusOpen, Classified
Product CodeKWP
Date Initiated2023-02-16
Product Quantity10 units

Reason for Recall

Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws

Firm Action

The firm issued an URGENT MEDICAL DEVICE RECALL NOTICE to its consignees on 02/16/2023 by UPS 2nd day mail. The letter explained the issue and the risk and requested the following actions be taken: Immediately identify, segregate, and quarantine affected products within your inventory. Pass this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Contact Medtronic to return affected product: 1-800-933-2635 to receive an return authorization (RGA) and coordinate inventory replacement/credit. Reference FA1315. The letter also includes specific patient recommendations.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide and the country of Canada.