Varian Medical Systems, Eclipse Treatment Planning System, with v client versions 7.3 (build 7.3.10), 8.0 (builds 7.5.x), 8.1 (builds 8.1.x) and 8.5( builds 8.2.x). For radiation therapy.

FDA Recall Z-1412-2010 · Posted 2010-04-20

Recall StatusTerminated
Product CodeMUJ
Date Initiated2009-12-09
Date Terminated2012-02-24
Product Quantity3177 unit

Reason for Recall

Potential for misdelivery of treatment-- When preparing a treatment plan involving a split, the device may calculate excessive or insufficient monitor units for the treatment plan.

Firm Action

Recall initiated on 12/09/2009. Varian issued a notification letter to its consignees, including a description and corrective action step. The firm is in the process of preparing a software fix for the issue.

Root Cause

Software design

Distribution Pattern

Nationwide distribution and Mexico.