Natus neoBLUE blanket LED Phototherapy system which includes the following items: 006224 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS
FDA Recall Z-1412-2015 · Posted 2015-04-10
Advertisement
Recalling FirmNatus Medical Incorporated
Recall StatusTerminated
Product CodeLBI
Date Initiated2015-03-16
Date Terminated2016-12-20
Product Quantity2611 units (1911 units in the US and 700 units outside the US)
Reason for Recall
neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the neoBLUE blanket system fiberoptic pads. These failures involve discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.
Firm Action
Natus sent the PRODUCT RELIABILITY ALERT neoBLUE blanket system performance along with an attached Technical Bulletin, Instructions for Use, and Service Manual to consignees via email during the week of March 16, 2015.
Natus emailed the Distribution Partner (DP) letter to all Natus International
Distribution Partners, requiring that they notify their customers (at Biomedical Department and Neonatal Intensive Care Unit) with copies of the Field Safety Notice, customer letter, the Technical Bulletin, Instructions for Use, and Service Manual.
The technical bulletin asks the customer to visually and tactilely inspect the fiber optic cable connector before use, and to call Natus if they notice a failing
pad. This procedure will prevent the complete degradation and therapy will not be compromised.
U.S. customers should contact Natus Technical Service at 888-496-2887 with any questions regarding this notification.
International customers should contact their local distributors or contact Natus Technical Service at (888) 496-2887 or [email protected].
Root Cause
Other
Distribution Pattern
distributed worldwide