Turon Impaction Fixture
FDA Recall Z-1412-2017 · Posted None
Advertisement
Recalling FirmEncore Medical, Lp
Recall StatusTerminated
Product CodeHWA
Date Initiated2017-02-09
Date Terminated2017-10-20
Product Quantity297 units
Reason for Recall
During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.
Firm Action
There are two field safety notices, one for Consignees who have surgeons that use the impaction fixtures (Verson 1) and one for Consignees who had previously indicated their surgeons do NOT use the impaction fixtures (Verson 2). The two recall notification letters were sent out on 2/9/17.
Root Cause
Device Design
Distribution Pattern
US, South Korea, Australia, Canada, United Kingdom/Ireland, Germany
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.