Leksell GammaPlan 11.1 Product Usage: Leksell GammaPlan is a computer-based system designed for Leksell Gamma Knife treatment planning.

FDA Recall Z-1412-2019 · Posted None

Recalling FirmElekta Instrument AB
Recall StatusOpen, Classified
Product CodeIWB
Date Initiated2018-06-27
Product Quantity52 devices

Reason for Recall

The margin tool in Leksell GammaPlan 11.1 systematically overestimates margin in certain areas of the volume.

Firm Action

Elekta initially notified customers on about 06/27/2018 via "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" LETTER. Instructions included to post the recall notice in a place accessible to all users, e.g. Instructions for Use, until the recall action is closed, advise the appropriate personnel working with the affected device of the recall and the content of the recall notification letter, and complete and return the Acknowledgement Form. Elekta updated their recall strategy on about 11/21/2018 and will be providing all affected customers with software upgrade 11.1.1. Electa service engineers will be pro-actively reaching out to customers to organize a date for installation on all affected devices. An updated recall notification was not sent to customers.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - US Nationwide distribution to GA, MD, NJ, PA, PR, WA. International distribution to Australia, Belgium, France, Germany, Hong Kong, Italy, Japan, Korea, Democratic People's Republic of, Netherlands, Romania, Russian Federation, Spain, Turkey, United Kingdom.