DeRoyal Sterile Custom Kits containing Cardinal Health Gowns. Sold as the following: Shoulder Pack 89-4751.08
FDA Recall Z-1412-2020 · Posted 2020-02-19
Advertisement
Recalling FirmDeroyal Industries, Inc. Lafollette
Recall StatusTerminated
Product CodeLRO
Date Initiated2020-02-04
Date Terminated2021-04-01
Product Quantity2100 units
Reason for Recall
Cardinal Health gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.
Firm Action
DeRoyal issued their recall on 1/26/2020 to the end user level by sending recall letters to its direct accounts. The customers were asked to identify the affected products and return the provided notice of return form back to DeRoyal indicating if replacement or credit is requested.
Root Cause
Environmental control
Distribution Pattern
US: Rhode Island and New Jersey Federal government agency: V.A Med Ctr East Orange
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.