MX8000IDT CT Scanner using version 2.5.5, 3.0, 3.1, 3.2, 3.2.2, or 3.2.4 software.

FDA Recall Z-1413-05 · Posted 2005-08-30

Recall StatusTerminated
Product CodeJAK
Date Initiated2005-03-11
Date Terminated2012-03-02

Reason for Recall

When two product applications called Flip and Secondary Capture are used together, it is possible for image orientation labels to be incorrect.

Firm Action

The firm sent recall/field correction letters dated 3/11/05.

Root Cause

Other

Distribution Pattern

The devices were distributed to consignees throughout the United States and to consigness located in Argentina, Austria, Bahama, Belarus, Belguim, Brazil, Canada, China, Chile, Columbia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, England, Finland, France, Greece, Germany, Hungary, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Lebanon, Libya, Mexico, Morocco, Netherlands, New South Wales, New Zealand, Norway, Panama, Phillipines, Portugal, Poland, Romania, Russia, South Africa, Singapore, Slovenia, Spain, Sri Lanka, Switzerland, Sweden, Taiwan, Thailand, Turkey and Venezuela.