Disetronic D-TRONplus Insulin Pump Battery Pack; Ref/Catalog no. 04697014001.

FDA Recall Z-1413-06 · Posted 2006-09-12

Recall StatusTerminated
Product CodeLZG
Date Initiated2006-07-13
Date Terminated2007-11-05
Product Quantity6,604 packs to U.S. and Canada.

Reason for Recall

The battery may turn the pump off without warning due to a design change in the battery.

Firm Action

Press release was issued on 7/13/06. Consignees, distributors and health care providers were issued a recall notice dated 7/13/06 instructing replacement of the power pack every two weeks until further notice.

Root Cause

Other

Distribution Pattern

Nationwide and Canada.