Cook Medical Tao Brush I.U.M.C. Endometrial Cytology/Histology Sampler (Tao Brush), 9.0Fr/26 cm, 3.5 cm, sterile, Cook Ob/Gyn, Spencer, IN; REF J-ES-090500-BL. Used to obtain endometrial samples for both cytology and histology. Intended for one-time use.
FDA Recall Z-1416-2009 · Posted 2009-06-16
Advertisement
Recalling FirmCook OB/Gyn, Inc.
Recall StatusTerminated
Product CodeHFE
Date Initiated2009-04-02
Date Terminated2009-12-03
Product Quantity15,396
Reason for Recall
There is an ink on the brush which has not been tested for use with the Tao Brush.
Firm Action
The consignee was notified by an Urgent: Recall Notice letter dated 4/2/09. The customer is to forward the notice to all personnel that need to be aware of it; pull from their stock and isolate any affected product; and complete the form. Upon receipt of the form, Cook Women's Health will issue a Return Goods Authorization number and the customer should return the affected products via FedEx using the Return Goods Authorization number.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.