BioGX SARS-CoV-2 Reagents for BD MAX System-IVD PCR test intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal specimens Catalog Number: 444213
FDA Recall Z-1416-2021 · Posted None
Advertisement
Recalling FirmBecton Dickinson & Co.
Recall StatusTerminated
Product CodeQJR
Date Initiated2021-03-24
Date Terminated2022-08-02
Product Quantity21,529 Kits
Reason for Recall
BioGX SARS-CoV-2 Reagents for BD MAX" System rehydration buffer tubes were not adequately sealed causing some tubes to be partially filled or empty upon use. A partially filled or empty rehydration buffer tube may lead to an Unresolved (UNR) result on the BD MAX" System
Firm Action
BD issued Urgent Medical Device Correction letter on March 24, 2021 FedEx tracked. Letter states reason for recall, health risk and action to take:
1. Inspect rehydration buffer tubes prior to use for the specific catalog and lot numbers listed below in Attachment A and destroy any products containing partially filled or empty tubes following your institutions process for destruction.
2. Share and Post this medical device recall notification within your facility network to ensure users inspect tubes prior to use.
3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement, if required.
4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via:
Actions Taken by BD:
1. Corrective actions have been initiated to prevent recurrence of the identified root cause.
2. BD will provide a product replacement for all destroyed inventory.
If you require further assistance, please contact: BD Contact US Contact Information Outside of US BD Customer/ Technical Support
800-638-8663
Monday Friday
7:00am and 7:00pm EST
For customers outside the US, contact the local BD representative or distributor.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.