LumiWave 1X4 Infrared Therapy Device, Catalog # LW1X4, BioCare Systems, Inc.
FDA Recall Z-1419-06 · Posted 2006-08-24
Advertisement
Recalling FirmBiocare Systems Inc.
Recall StatusTerminated
Product CodeILY
Date Initiated2006-06-14
Date Terminated2011-01-03
Product Quantity175 units
Reason for Recall
Electrical Safety Hazard-when pulling on the cord of the AC adapter, to remove it from an outlet, the prongs on the AC adapter, may detach from the unit, damaging it and exposing the user to an electrical safety hazard
Firm Action
Consignees were notified by telephone beginning 06/08/2006, followed by a letter on 06/23/2006.
Root Cause
Other
Distribution Pattern
Nationwide distribution --- including the states of CA, CO, IL, MA, MD, MI, NC, FL, PA, TX, UT, WA.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.