Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

FDA Recall Z-1420-2018 · Posted None

Recalling FirmAutochair
Recall StatusOpen, Classified
Product CodeFSA
Date Initiated2018-03-07

Reason for Recall

Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.

Firm Action

Autochair will issue a Retrofit Failsafe Kit. In the meantime, please quarantine current stock of SmartBase products until the failsafe device has been fitted and tested.

Root Cause

Component design/selection

Distribution Pattern

PA, and worldwide