Terminated

IMMULITE 2000 Progesterone; Catalog Number: L2KPW2 L2KPW6; Siemens Material Number: 10381181, 10381170 Product Usage: For in vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers - for the quantitative measurement of progesterone in serum as an aid in the diagnosis and treatment of disorders of the ovaries or placenta.

FDA Recall Z-1421-2017 · Posted None

Recall StatusTerminated
Product CodeJLS
Date Initiated2017-01-04
Date Terminated2019-07-18
Product Quantity11963

Reason for Recall

DHEA-S causes falsely elevated progesterone results.

Firm Action

Siemens sent an Urgent Medical Device Correction (UMDC) letter dated January 4, 2017 to all affected Siemens Healthcare Diagnostics customers in the United States dated . An Urgent Field Safety Notice (UFSN) was sent out for distribution to all customers outside the United States on January 4, 2017. These letters identified the affected product problem and actions to be taken. For questions contact your local Siemens Customer Care Center or your local Siemens Technical support.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution including Puerto Rico

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.