STA-R blood coagulation analyzer. Catalog Number 57160.
FDA Recall Z-1422-05 · Posted 2005-08-30
Advertisement
Recalling FirmDiagnostica Stago, Inc.
Recall StatusTerminated
Product CodeJPA
Date Initiated2005-06-20
Date Terminated2005-08-30
Product Quantity215 units
Reason for Recall
Concerning the STA-R blood coagulation analyzer, there is a possible failure mode associated with patient samples receiving prolonged exposure time to reagents during the cuvette roll change process.
Firm Action
Alert on the problem was sent out on 6/20/2005 to all customers alerting the problem and instructions on how to prevent the problem. The units are not be to returned.
Root Cause
Other
Distribution Pattern
Units were distributed to clinics and hospitals nationwide.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.