Terminated

Cobe CML Duo Flat Sheet Membrane Oxygenator, sterile, (Custom Heart/Lung Packs), Catalog numbers: 067100001, 078002103, 078005802, 078159025, 078201013, 078212009, 078310006, 078332022, 078348014, 078428010, 078688007, 078869003. COBE Cardiovascular, Inc..

FDA Recall Z-1426-06 · Posted 2006-08-24

Recall StatusTerminated
Product CodeDTZ
Date Initiated2006-06-26
Date Terminated2011-08-08
Product Quantity580 units (Domestic)

Reason for Recall

Cooling and rewarming time is extended. Efficiency of the heat exchangers is reduced because many of the folds that are critical to heat exchange on the blood side of the units were blocked with a metalic material, thus decreasing the surface area needed for proper heat transfer.

Firm Action

Consignees were notified by letter on 06/26/2006.

Root Cause

Other

Distribution Pattern

Worldwide distribution --- USA including VA facilities in CO, GA, MO and countries of Australia, Canada, Italy, and Japan