Thyroidectomy Indication for the da Vinci Surgical Systems; Product Usage: da Vinci Thyroidectomy offers a minimally invasive surgery for both partial and total thyroidectomy procedures, including endoscopic and video assisted procedure that are preferable to conventional (open) procedures that utilize a cervical incision because they leave no visible neck scar.

FDA Recall Z-1426-2013 · Posted 2013-05-29

Recall StatusTerminated
Product CodeNAY
Date Initiated2011-10-13
Date Terminated2013-05-30
Product Quantity2,423 consignee

Reason for Recall

Promotional literature for use of the da Vinci system for thyroidectomy, which had not been approved with a change to the 510(k).

Firm Action

Intuitive sent a notification letter on October 17, 2011 by email to all consigness whose email address were on file. Also a copy of the letter was sent by registered mail on February 8, 2012 to all surgeons whose addresses were on file. The letters informed consignees that Intuitive Surgical were in the process of reviewing labeling for the da Vinci System in regard to use for thyroidectomy procedures. During this period, Intuitive will not provide or facilitate any promotional, training, case observations, proctoring or in case procedural support for use of the da Vinci System in conjunction with thyroidectomy procedures. The letters recommend consulting the institution policy regarding use of approved or cleared medical devices for procedures that are not specified in the labeled indication for use. Questions may be directed to customer service at 1-877-408-3872.

Root Cause

No Marketing Application

Distribution Pattern

USA Nationwide distribution