Recall StatusOpen, Classified
Date Initiated2022-05-27
Product Quantity667 devices
Reason for Recall
The base of the anesthesia system can have a crack and could potentially fracture resulting in a tip or overbalance of the anesthesia system. This can result in potential injury if it falls on a person.
Firm Action
The recalling firm issued letters dated 5/27/2022 via traceable means on 5/27/2022 which were addressed to the Director of Respiratory, Chief of Anesthesia, Health Care Administrator/Risk Manager, and Director of Biomedical/Clinical Engineering. The letter is flagged that the consignee needs to ensure all potential users in the facility are made aware of this safety notification and recommended actions. The notification informs the consignees of the affected devices, the safety issue, and actions to be taken. The actions notified the consignee they can continue to use the anesthesia system by following instructions listed in the letter. The instructions were to inspect the anesthesia system for cracks at the rear of the base as depicted in the diagram and photograph, and if no cracks are observed, they can continue to use the anesthesia system. If a crack at the rear is observed, they can continue to use the system under the following conditions: Limit the movement of the system using care on uneven flooring/threshold and ensure the floor is clear of obstacles (cables, power cords, etc.); and do not exceed the recommended weight limit for equipment mounted or supported by the anesthesia system. A Medical Device Notification Acknowledgement Response form was to be completed and returned within 30 days of receipt to acknowledge receipt and understanding of the notification. The letter informs the consignee that the recalling firm will inspect all affected devices at no cost and that a firm representative will contact the consignee to arrange for the correction.
Root Cause
Process control
Distribution Pattern
Distribution was made to AZ, CA, CO, CT, FL, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, MT, NC, NE, NV, NY, OH, PA, SC, TN, TX, VA, and WV. There was also military distribution but no government distribution.
Foreign distribution was made to Australia, Canada, China, Estonia, Finland, France, Germany, Greece, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Korea, Malaysia, New Zealand, Philippines, Romania, Russia, Saudi Arabia, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900, Aespire View, Avance, Avance CS2, Aisys, Aisys CS2Datex-Ohmeda, Inc. · 2017-11-03
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.