Roche/Hitachi 747 - 100 clinical chemistry analyzer; catalog number 04009223680.
FDA Recall Z-1428-04 · Posted 2004-09-02
Recalling FirmRoche Diagnostics Corp.
Recall StatusTerminated
Product CodeJJE
Date Initiated2004-08-06
Date Terminated2006-06-26
Product Quantity254
Reason for Recall
A software defect will allow the reporting out of believable, but clinically significant, erroneous results if the analyzer is in operation and a channel has been masked or not requested over a period of time.
Firm Action
Recall letters dated 8/6/04 were sent to each customer advising them of the problem and giving instructions on how to avoid it.
Root Cause
Other
Distribution Pattern
United States.