Roche/Hitachi 747 - 100 clinical chemistry analyzer; catalog number 04009223680.

FDA Recall Z-1428-04 · Posted 2004-09-02

Recall StatusTerminated
Product CodeJJE
Date Initiated2004-08-06
Date Terminated2006-06-26
Product Quantity254

Reason for Recall

A software defect will allow the reporting out of believable, but clinically significant, erroneous results if the analyzer is in operation and a channel has been masked or not requested over a period of time.

Firm Action

Recall letters dated 8/6/04 were sent to each customer advising them of the problem and giving instructions on how to avoid it.

Root Cause

Other

Distribution Pattern

United States.