Roche/Hitachi Modular Analytical D Module clinical chemistry analyzer; catalog number 03739023001.

FDA Recall Z-1429-04 · Posted 2004-09-02

Recall StatusTerminated
Product CodeJJE
Date Initiated2004-08-06
Date Terminated2006-06-26
Product Quantity138

Reason for Recall

A software defect will allow the reporting out of believable, but clinically significant, erroneous results under certain conditions.

Firm Action

Recall letters dated 8/6/04 were sent to each customer advising them of the problem and giving instructions on how to avoid it.

Root Cause

Other

Distribution Pattern

United States.