Centurion kits: a) DENTAL INSTRUMENT KIT, Product Code DH760; b) STERILE COTTON DENTAL ROLLS 6/PK, Product Code 9125ST6; c) DENTAL INSTRUMENT KIT, Product Code DH760

FDA Recall Z-1431-2024 · Posted 2024-03-29

Recall StatusOpen, Classified
Product CodeOFX
Date Initiated2024-02-23
Product Quantity2310 units

Reason for Recall

Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.

Firm Action

Medline Industries, LP, issued a Medical Device Recall notice to its consignees on 02/27/2024 via email and USPS First Class mail. The notice explained the problem with the devices, risk, and requested that the products be destroyed at the consignee. Distributors were directed to notify their customers.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Panama, Canada.