Terminated

COULTER LH 500 Hematology Analyzer part numbers: 178832, 178833, and 178834 software versions 1A and 1A2. The LH 500 Analyzer is a quantitative, automated hematology analyzer and leukocyte differential cell counter For In Vitro Diagnostic Use in clinical laboratories. The LH %500 Analyzer also provides a semi-automated reticulocyte analysis.

FDA Recall Z-1433-04 · Posted 2004-09-02

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeGKZ
Date Initiated2004-08-12
Date Terminated2012-04-26
Product Quantity222

Reason for Recall

Beckman Coulter has confirmed an issue associated with the predilute mode of the COULTER LH 500 Hematology Analyzer. The workstation may display, transmit, and / or print an erroneous result or an incorrect dilution factor with a predilute sample result.

Firm Action

Customers were notified via U.S. mail by a Product Corrective Action Letter (PCA) dated August 12, 2004. The letter includes a description of what conditions need to be met for the error to occur and the proper steps to prevent the issue from occurring.

Root Cause

Other

Distribution Pattern

Nationwide & Canada

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.