Roche/Hitachi ALT (ALAT/GPT) Reagent Kit for use on the Roche/Hitachi MODULAR ANALYTICS P Module; Catalog number 1877470 (11877470216).
FDA Recall Z-1434-04 · Posted 2004-09-02
Advertisement
Recalling FirmRoche Diagnostics Corp.
Recall StatusTerminated
Product CodeCKH
Date Initiated2004-08-20
Date Terminated2004-11-05
Product Quantity728 non-expired kits.
Reason for Recall
Rotation of the instrument reagent compartment can cause a bubble to form in the neck of the reagent bottle causing premature liquid level detection and short sampling of the reagent, resulting in falsely low or zero patient results.
Firm Action
Consignees were mailed recall notices dated 8/20/04 and were instructed to either reduce the initial reagent bottle volume prior to use, or to request replacement product.
Root Cause
Other
Distribution Pattern
Nationwide.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.