ConMed DetachaTip Multi-Use Laparoscopic Instrument, "Endoweave Multiple Use Grasper", 5 mm x 33 cm, REF/Product code 1-1028. -------- The product is distributed sterile in heat-sealed tray and labeled with an expiration date that pertains only to the sterility.

FDA Recall Z-1434-06 · Posted 2006-08-24

Recalling FirmConmed Corporation
Recall StatusTerminated
Product CodeGEI
Date Initiated2006-03-27
Date Terminated2012-04-27
Product Quantity3496 units

Reason for Recall

Grasper jaws breaking at the junction of the jaw and tube during laparoscopic procedures.

Firm Action

On 3/27/06, all domestic customers were sent an "URGENT DEVICE RECALL" notice, dated 3/27/06, by USPS Priority Mail (Delivery Confirmation). International customers were notified by facsimile and/or e-mail as well as by mail.

Root Cause

Other

Distribution Pattern

Worldwide