Orthofix US XCaliber Ankle Complete Kit, Sterile, 99-91647. manufactured by Orthofix SRL, Via delle Nazioni, 9, 27012 Bussolengo (VR) Italy. Intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation and other bone conditions amenable to treatment by use of the external fixation modality.

FDA Recall Z-1435-2010 · Posted 2010-04-28

Recalling FirmOrthofix Srl
Recall StatusTerminated
Product CodeJDW
Date Initiated2010-02-22
Date Terminated2010-05-05
Product Quantity25

Reason for Recall

During standard quality controls, firm identified a potential for breakage of the XCaliber clamp when locking the cam.

Firm Action

Orthofix issued an "Urgent: Field Safety Notice" dated February 22, 2010. Consignees were informed of the affected product and instructed to cease use and to return the product to the firm.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution -- TX, ND, NY, FL and CA.