IMPAX CV Results Manager/Results Manager Administration Tool The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting. The suite is comprised of modular software that runs on industry standard personal computers and servers, utilizing Windows 2000/2003/XP operating systems. Image data storage/display utilizes DICO< (Digital Imaging and Communications in Medicine) data primarily from independent medical devices.
FDA Recall Z-1435-2011 · Posted 2011-02-25
Advertisement
Recalling FirmAGFA Corp.
Recall StatusTerminated
Product CodeLLZ
Date Initiated2009-04-15
Date Terminated2011-10-13
Product Quantity6 devices
Reason for Recall
User could modify a measurement's units and that change would not be reflected in echocardiography formulas used within the Adult echocardiography RM.
Firm Action
The firm, Agfa Healthcare, sent individual letters dating from April 15, 2009 to August 26, 2009, to the customers. The letters described the product, problem and actions to be taken. Agfa provided the affected customer with a complete list of their impacted patients and the measurements. Agfa also scheduled unit corrections with their customers and received acknowledgement of the letters at the time of scheduling the unit corrections. Service Bulletin DIS026.09E was issued January 2009 to inform users that Agfa will no longer allow customers to request unit changes for measurements that have formulas. Note: Technical Support ran diagnostic scripts for each customer and documented in GSC cases for each consignee/customer.
If you have any questions, contact Agfa Healthcare at (949) 830-4959.
Root Cause
Other
Distribution Pattern
Nationwide distribution: USA including states of: IL, MT, NY, OR and SC.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.