D826 Accessory kit for single tcpO2 Electrodes for use with the TCM400 Transcutaneous Monitor (Cutaneous Oxygen Monitor)

FDA Recall Z-1436-06 · Posted 2006-08-31

Recall StatusTerminated
Product CodeKLK
Date Initiated2005-11-01
Date Terminated2012-04-25
Product Quantity3,323-USA, 1,197-Foreign

Reason for Recall

Excessive Drift-the membrane units of the device cause the electrode to exceed performance standards for drift in the first two or more calibrations

Firm Action

The firm decided to initiate a recall on 9/8/2005 and the recall letter was issued in November, 2005. The letter informs the customers of this problem and asks that each customer inspect their current stock for the presence of the recalled lot codes and discard any such product found. The letter is accompanied by a recall response form which the customer can use to request additional replacement ellectrodes as necessary.

Root Cause

Other

Distribution Pattern

Worldwide distribution --- including the states of PA, IL, TX, NY, FL, KS, NE, AL, MI, LA, WI, MN, NJ, CA, NH, MD, OH, MO, UT, IN, GA, WA, MA, KY, NC, IA, OK, SD, CO, AR, SC, ME, NV, AZ, RI, AK, MS, VA, WV, MT, TN, IN & NM. Country List pending

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →