Terminated

DirectInject consists of a sterile dual paste system which is calcium phosphate based. Upon injection through the Mixer-Cannula, the two pastes form a cement paste which is moldable. The injected cement paste will harden under normal body conditions to form hydroxyapatite.

FDA Recall Z-1436-2021 · Posted None

Recall StatusTerminated
Product CodeGXP
Date Initiated2021-04-01
Date Terminated2022-12-19
Product Quantity321 units

Reason for Recall

Inability for the user to inject the paste from the syringe into the target location.

Firm Action

On 04/01/2021, Stryker issued an Urgent Medical Device Removal notice via letter informing customers that users have reported the inability to push the paste through the Mixer-Cannula which prohibits the injection into a bone void.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide distribution and the countries of Germany, Italy, France, Australia, Japan.