Diopsys Inc., Enfant Pediatric VEP Vision Testing System, Input: 120 Vac, 60Hz/2.5A IEC 60601-1 Class I, Type BF.
FDA Recall Z-1437-2010 · Posted 2010-04-26
Advertisement
Recalling FirmDiopsys, Inc.
Recall StatusTerminated
Product CodeGWE
Date Initiated2007-04-02
Date Terminated2010-04-27
Product Quantity175 units
Reason for Recall
The computer based system may lose test counts when the time in the device was changed during a changeover from standard time to daylight savings time in March 2007.
Firm Action
Field Service personnel were sent an e-mail notification on 4/2/2007 which included a TZedit file to be downloaded. E-mails instructed reps to install the patch on their customers devices immediately. Field correction was completed 7/31/2007.
Root Cause
Software design
Distribution Pattern
Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.