LKT200 Perfusion Circuit, part of Lifeport Kidney Perfusion Transporter KTR - Product Usage: The circuit is an accessory to LifePort Kidney Perfusion Transporter (KTR) Isolated kidney perfusion and transport system, which is intended to be used for the continuous hypothermic machine perfusion of kidneys for the preservation, transportation and eventual transplantation into a recipient.

FDA Recall Z-1437-2020 · Posted None

Recall StatusTerminated
Product CodeKDN
Date Initiated2020-01-30
Date Terminated2021-09-30
Product Quantity2274

Reason for Recall

Observed leakage at the sample port assembly which could potentially lead to loss of perfusion solution needed to maintain continuous machine perfusion.

Firm Action

Letters were sent 01/30/2020 to all customers advising cease usage and quarantine all LKT200 units and return to firm.

Root Cause

Reprocessing Controls

Distribution Pattern

Distributed nationwide (32 states) and to Canada.