Terminated

adult defibrillation, cardioversion, pacing, monitoring Disposable Electrodes, Model P-211-M1 and P-214-M1. Product Usage: Used by trained professionals in hospitals, doctors offices, and Emergency Medical Services for adult low-energy defibrillation, transcutaneous pacing, cardioversion and monitoring.

FDA Recall Z-1438-2013 · Posted 2013-05-30

Recalling FirmTZ Medical Inc.
Recall StatusTerminated
Product CodeDRO
Date Initiated2013-04-30
Date Terminated2013-09-05
Product QuantityUS: 10,410 pieces; Outside US: 100 pieces

Reason for Recall

TZ Medical Inc is recalling its P-211-M1 and P-214-M1 Medtronic / Physio Control Compatible transparent defibrillation electrodes because it may contain a defective white wire on the connector assembly. If a failure were to occur, the end user would be able to identify the failure by inspecting the white wire 1/4 inch from the Gray Medtronic / Physio adapter. The consignees will notice discolor

Firm Action

TZ Medical Inc. started notifying their customers by telephone on April 30, 2013. The firm followed up with an Urgent: Medical Device Recall notification letter dated May 14, 2013, to their consignees. TZ Medical Inc., Representatives have will contact all affected customers and arrange for the replacement of the effected product in stock with inspected product. For any questions about this recall, please fax to 503-639-0239.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - USA (nationwide) and Chile.