Terminated

bellavista 1000 ventilator, Catalog number 301.100.030. bellavista is an electronically controlled pneumatic ventilation system with an integrated air compressing system. bellavista uses room air and high-pressure oxygen. Air enters through a fresh gas intake port and is compressed together with the oxygen by the blower. Oxygen enters through a high pressure inlet. An electronic mixer valve provides for the operator-set concentration. Gas is supplied to the patient via the microprocessor controlled inspiratory valve.

FDA Recall Z-1439-2020 · Posted None

Recalling FirmVyaire Medical
Recall StatusTerminated
Product CodeCBK
Date Initiated2019-11-25
Date Terminated2023-08-08
Product Quantity116

Reason for Recall

The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority alarm, presence of a 'no alarm' condition, or presence of non-responsive touch screen.

Firm Action

Consignees were sent an Urgent Medical Device Recall letter beginning 11/25/2019 via overnight delivery. The letter identified affected devices, provided a description of the issue, and provided actions to be taken by distributors and end-users. A response form was asked to be returned. Questions can be directed to Vyaire's Technical Support Department at [email protected].

Root Cause

Software design

Distribution Pattern

US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, MD, MI, PA, TX, WA and Puerto Rico.