Centurion Curette kits labeled as: a) STERILE BUCK EAR CURETTE "00" (25697), Product Code 66745; b) STERILE "1" EAR CURETTE (505629), Product Code 67050; c) STERILE FOX DERMAL CURETTE, 5MM(32923), Product Code 67430; d) STERILE EXCAVATOR CURETTE (P/S), Product Code I67970; e) ST BUCK EAR CURETTE BLNT STR SZ 00, Product Code I68040; f) STERILE BRUNS CURETTE #4 (BC400S), Product Code I68340; g) 6MM FOX DERMAL CURETTE STERILE, Product Code I68380; h) STERILE KEVORKIAN CURETTE 4MM (KEC4S), Product Code I68470; i) ST ENDOCERVICAL BIOPSY CURETTE 2MM, Product Code I68480

FDA Recall Z-1439-2024 · Posted 2024-03-29

Recall StatusOpen, Classified
Product CodeHTF
Date Initiated2024-02-23
Product Quantity20614 units

Reason for Recall

Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.

Firm Action

Medline Industries, LP, issued a Medical Device Recall notice to its consignees on 02/27/2024 via email and USPS First Class mail. The notice explained the problem with the devices, risk, and requested that the products be destroyed at the consignee. Distributors were directed to notify their customers.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Panama, Canada.