Zeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI Pentero C, Product Numbers: 302583-9001-000 (PENTERO C), and 305961-0000-00 (Ceiling Mount for PENTERO C), Product Usage: A surgical microscope and accessories is an AC-powered device intended for use during surgery to provide a magnified view of the surgical field.

FDA Recall Z-1440-2013 · Posted 2013-05-31

Recall StatusTerminated
Product CodeEPT
Date Initiated2013-04-18
Date Terminated2013-06-26
Product Quantity6 units

Reason for Recall

Carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical microscope system. During a routine inspection, one microscope installation was found to have 3 (three) broken screws out of 6 (six) in the lower section of the ceiling mount.

Firm Action

Carl Zeiss Meditec, Inc sent a Customer Notification letter to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that firm's sales representative will be scheduling an on-site visit, informing them of the affected product and providing instructions on the recall. For question call 925-580-5371.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide Distribution.