Terminated

Varian Medical Systems On-Board Imager, Versions 1.0.15 and 1.2.05 Imaging Accessory to Medical Linear Accelerator. The On-Board Imager Device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial landmarks.

FDA Recall Z-1442-2008 · Posted 2008-08-15

Recall StatusTerminated
Product CodeIYE
Date Initiated2008-01-03
Date Terminated2010-11-27
Product Quantity291 units

Reason for Recall

If used with a third party radiation therapy treatment planning software system, mistreatment may occur because of a misalignment.

Firm Action

The firm has upgraded all user sites through normal course of service and installation activities. Notification letters were distributed starting January 03, 2008.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution -- USA including states of GA, MI, NY, MN, TX, TN, CA, MD, KS, AL, KY, CT, WI, LA, NC, NJ, PA, IL, RI, FL, NE, AK, VA, WV, NV, MO, TN, OR, IN, MA, CT, AL, AZ, NH, ID, OH, SC, CO, CT, ND, WA, and UT, and countries of Sweden, Switzerland, Germany, Denmark, Japan, UK, France, Italy, Norway, Taiwan, New Zealand, Canada, Poland, Australia, Colombia, China, Spain, Belgium, Russian Federation, India, Iceland.