Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Instruction for Use for Symphony Double Pumping Kit; Medela, Inc., P.O. Box 660, 1101 Corporate Drive, McHenry, IL 60051-0660; the instructions were packaged with the following non-sterile single user accessory kits for use with Symphony breast pumps: a) Model 67023, Harmony to Symphony Conversion Kit; b) Model 67091, Lactina to Symphony Conversion Retail Kit; c) Model 67099, Symphony Retail Kit; d) Model 67099-06, Symphony Retail Kit - case of 6 kits; e) Model 67099NA, Symphony Retail Kit - case of 6 kits for national accounts;

FDA Recall Z-1442-2010 · Posted 2010-04-27

Recalling FirmMedela Inc
Recall StatusTerminated
Product CodeHGX
Date Initiated2010-03-15
Date Terminated2010-12-10
Product Quantity13,834 kits

Reason for Recall

The non-sterile devices were erroneously packaged with instructions stating the devices are sterile and do not require cleaning before first use.

Firm Action

An "IMPORTANT:DEVICE RECALL" letter dated March 12, 2010 to the the customers on March 15, 2010, via certified mail, return receipt requested. The letters were tailored to each customer listing the affected shipments sent to them, and informing them of the labeling insert problem. The customers were requested to cease distribution of the affected kits and return them to the firm (Medela) for replacement, and if applicable, to notify their retail customers. The customers were instructed to call Medela Customer Service at 877-320-2301 for assistance in processing returns or for further clarification on the recall.

Root Cause

Labeling Change Control

Distribution Pattern

Nationwide distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑