ProTrack Microcatheter; Product Usage: The ProTrack Microcatheter is a sterile, single use device intended to be used as a standard infusion wire or as an exchange guide for exchanging guidewires.

FDA Recall Z-1442-2015 · Posted 2015-04-13

Recalling FirmBaylis Medical Corp *
Recall StatusTerminated
Product CodeDQX
Date Initiated2015-03-30
Date Terminated2016-11-17
Product Quantity58 units

Reason for Recall

Baylis Medical Company is recalling two lots of ProTrack Microcatheter due to the possibility that the catheter may have circumferential defects along its shaft.

Firm Action

A recall notification, marked "URGENT" will be sent by Baylis Medical via overnight courier to all customers affected by the recall with a fax-back form. The form is to be faxed back to Baylis Medical confirming receipt of the recall notice. As per instructions on the fax-back form, affected devices are requested to be returned. Baylis Medical will replace the returned affacted catheters with a new catheter.

Root Cause

Process control

Distribution Pattern

US Nationwide Distribution - US Nationwide in the states of MO, MA, OH, CA, AR, MA, FL, GA, CO, TX, KY, NY, MI, WI, OR