NanoSight NS300
FDA Recall Z-1442-2022 · Posted None
Advertisement
Recalling FirmMALVERN PANALYTICAL LTD
Recall StatusOpen, Classified
Product CodeREN
Date Initiated2022-04-07
Product Quantity507 units
Reason for Recall
Product found to not comply to Class 1 Laser safety requirement.
Firm Action
Customers are being made aware of this issue. All affected customers will be provided with a remediation kit with Class 3R labels and
manual addenda with revised safety instructions.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US Nationwide - Worldwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.