Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

NanoSight NS300

FDA Recall Z-1442-2022 · Posted None

Recall StatusOpen, Classified
Product CodeREN
Date Initiated2022-04-07
Product Quantity507 units

Reason for Recall

Product found to not comply to Class 1 Laser safety requirement.

Firm Action

Customers are being made aware of this issue. All affected customers will be provided with a remediation kit with Class 3R labels and manual addenda with revised safety instructions.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

US Nationwide - Worldwide Distribution

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑