Philips Medical Systems iSite PACS (Picture Archiving and Communications Systems) version 3.6.28.0 - Radiology Image management system.

FDA Recall Z-1443-2008 · Posted 2008-08-16

Recall StatusTerminated
Product CodeLLZ
Date Initiated2008-01-29
Date Terminated2008-09-25
Product Quantity3 units

Reason for Recall

When using the Freehand Region of Interest (ROI) tool, errors may occur in area calculation yielding incorrect results. The firm discovered that a descrepancy exists in the measurement provided by the tool.

Firm Action

The firm issued product notification letter February 20, 2008, to its consignees. The notification alerts customers to the behavior, requires them to post the notification and notify users not to use the Freehand ROI feature until they are upgraded to a corrected software version. Philips will deliver a software upgrade to correct the problem.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution --- USA including states of MN, CO, and PA, and country of Germany.