Delphi IVantage Volumetric Ambulatory Infusion Pump/System; Ref nos. 12864051FRE, Rev. 8 and 12864051DEL, Rev. 8.

FDA Recall Z-1445-06 · Posted 2006-08-26

Recall StatusTerminated
Product CodeFRN
Date Initiated2006-03-29
Date Terminated2006-10-16
Product Quantity22 US and 147 international.

Reason for Recall

Potential for under-infusion without alarm; cassette rollers stop moving , but the pump shaft continues rotating without alarm.

Firm Action

Distributors were notified via letter dated 3/29/05 instructing them to retrieve and return the product.

Root Cause

Other

Distribution Pattern

Florida, Austria, Denmark, Finland, Germany, Norway, Sweden, Switzerland and United Kingdom.