Olympus Resection Sheath, 28 Fr.; Model No. A22043A. Resection sheath for urologic applications.

FDA Recall Z-1445-2026 · Posted 2026-02-20

Recall StatusOpen, Classified
Product CodeFJL
Date Initiated2025-11-11
Product Quantity118 units

Reason for Recall

Complaints of the ceramic tip of the resection sheath breaking have been received.

Firm Action

An URGENT MEDICAL DEVICE CORRECTIVE ACTION notice was sent to consignees 1/6/2026. Consignees are instructed to ensure all personnel are familiar with the contents of the notification. Users are to adhere to the Inspection and Testing and Caution sections of the IFU for both the resection/inner sheath, as well as the Inspection Before Each Use section in the Resectoscope System Manual. Additional instructions include having an extra resection sheath and inspect during and after procedure for device fragments. Consignees are to retain a copy of the notification with the IFU. Consignees with any questions may contact Cynthia Ow at [email protected] or by phone at 947-999-3203.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide Distribution.