Centurion Manual surgical kits labeled as: a) 15ML SAMPLING TUBE WITH CAP, Product Code 15MLVLCP; b) STERILE 9X12 RECLOSABLE BAG, Product Code MG912ST
FDA Recall Z-1448-2024 · Posted 2024-03-29
Advertisement
Recalling FirmMEDLINE INDUSTRIES, LP - Northfield
Recall StatusOpen, Classified
Product CodeFMH
Date Initiated2024-02-23
Product Quantity1700 units
Reason for Recall
Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.
Firm Action
Medline Industries, LP, issued a Medical Device Recall notice to its consignees on 02/27/2024 via email and USPS First Class mail. The notice explained the problem with the devices, risk, and requested that the products be destroyed at the consignee. Distributors were directed to notify their customers.
Root Cause
Process control
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Panama, Canada.