Description/Part: Simplify Disc Size 1, Ht 5/SM-5; Simplify Disc Size 1, Ht 6/SM-6; Simplify Disc Size 2, Ht 4/MD-4; Simplify Disc Size 2, Ht 5/MD-5; Simplify Disc Size 2, Ht 5, 5/MD-5L; Simplify Disc Size 2, Ht 6/MD-6; Simplify Disc Size 2, Ht 6, 5/MD-6L; Simplify Disc Size 3, Ht 5/LG-5; Simplify Disc Size 3, Ht 5, 5/LG-5L; Simplify Disc Size 3, Ht 6/LG-6; Simplify Disc Size 3, Ht 6, 5/LG-6L
FDA Recall Z-1453-2022 · Posted None
Advertisement
Recalling FirmNuVasive Inc
Recall StatusOpen, Classified
Product CodeMJO
Date Initiated2022-06-20
Product Quantity1,963
Reason for Recall
Cervical artificial disc Unique Device Identifier (UDI) barcodes contain the incorrect Global Trade Item Number (GTIN) on the Patient Label when scanned and may also contain the incorrect GTIN human readable text, which could lead to selecting an incorrectly sized unit.
Firm Action
On 06/20/22, Field Safety notices were delivered in person, emailed, or mailed to hospital and distributor consignees who were informed of the following:
1) Complete and return the consignee confirmation form.
2) Return of affected devices is not required. However, if you would like these items
exchanged, please contact your firm representative to request a Return Material
Authorization number (RMA).
3) This recall notice should be shared with anyone who needs to be aware within your
organization or shared with any organization where the potentially affected products have been transferred.
Customers with questions, complaints, or who want to report adverse effects can contact the firm at [email protected]
Root Cause
Labeling Change Control
Distribution Pattern
US: CA, WA, TX, IN, MO, MI, FL, VA, OR. OUS: DE, NZ, SG, UK
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.