Terminated

Pioneer Surgical brand Quantum Spinal Rod System Locking Cap; Catalog No 10-LCAP.

FDA Recall Z-1457-05 · Posted 2005-08-31

Recall StatusTerminated
Product CodeMVH
Date Initiated2005-07-29
Date Terminated2005-10-25
Product Quantity5069

Reason for Recall

The product has a higher than normal locking torque, which may cause the saddle to malfunction or to break interoperatively.

Firm Action

Distributors were notified via letter dated July 29, 2005 to retrieve all caps from inventory and point of use locations and to return them to Pioneer.

Root Cause

Other

Distribution Pattern

Nationwide.