Merge Hemo software. Product Usage: Merge Hemo is a hemodynamic monitoring system that records and displays physiological data.
FDA Recall Z-1457-2017 · Posted None
Advertisement
Recalling FirmMerge Healthcare, Inc.
Recall StatusTerminated
Product CodeDQK
Date Initiated2016-04-04
Date Terminated2019-08-05
Product Quantity361 sites potentially have the affected versions
Reason for Recall
In some instances, the system will lock tabs within a study, even when a second user does not have the study open on a different workstation.
Firm Action
The recalling firm issued a letter dated March 28, 2016 via e-mail. If no email address was available, the letters were issued via certified mail. Non-responders were issued letters on July 11, 2016. The letter identified the affected product, problem and actions to be taken. For question send an email to [email protected].
Root Cause
Software design
Distribution Pattern
US Nationwide Distribution.