High-Definition 120 MultiLeaf Collimator, Medical charged-particle radiation therapy system; Software Version 7.0; Product Code: H56; Accessory to Medical Linear Accelerator.

FDA Recall Z-1459-2008 · Posted 2008-08-02

Recall StatusTerminated
Product CodeIYE
Date Initiated2008-01-30
Date Terminated2010-08-18
Product Quantity3 units subject to correction

Reason for Recall

Dose Calculation Error: Software anomaly may result in failure of a MLC leaf to reach planned position, potentially resulting in misadministration of dose to a patient.

Firm Action

Consignees were notified by a Medical Device Notification letter sent on 01/29/2008. The letter advised users to cease use of the device until a corrected version of the software is available. A representative will contact users to upgrade the application once the revised software has been validated and released for distribution.

Root Cause

Software design

Distribution Pattern

Nationwide Distribution including states of NC, FL and CA.