Medline ReNewal Reprocessed Livewire Steerable Electrophysiology Catheters, Medline Item Numbers: 401582R, 401904R, 401904R, 401905R, 401933R, and 401991R

FDA Recall Z-1460-2025 · Posted 2025-03-26

Recall StatusOpen, Classified
Product CodeNLH
Date Initiated2025-02-06
Product Quantity74 units

Reason for Recall

Devices may have higher than expected amounts of bacterial endotoxin.

Firm Action

An IMMEDIATE ACTION REQUIRED notification letter dated 2/6/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item numbers and the affected lot numbers listed on the attachment. Quarantine all affected product. 2. Please return the completed enclosed response form listing the quantity of affected product on hand. Even if you do not have any affected product, please complete and return the form, as indicated on the response form. When we receive your completed response form you will be sent FedEx return labels to return the product back to Medline ReNewal (if applicable). Your account will receive credit or replacement when the returned product is received. 3. If you have transferred possession of this product to another individual, to another department or another location in your organization you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline ReNewal. If you have any questions, please contact 866-359-1704.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide distribution in the states of AR, AZ, CA, CO, IA, IL, KS, OR, TN, TX, WA, and WI.