Brilliance CT 16, Brilliance CT 64, Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology, Brilliance iCT, Ingenuity CT, Ingenuity Core, Ingenuity Core 128 and Ingenuity Flex scanners are Computed Tomography X-ray systems.

FDA Recall Z-1461-2014 · Posted 2014-04-16

Recall StatusTerminated
Product CodeJAK
Date Initiated2014-03-27
Date Terminated2016-02-02
Product Quantity135

Reason for Recall

Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.

Firm Action

A Field Safety Notice was sent to all US consignees via certified mail. Market representatives will distribute the letter internationally. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution - USA including the states of CA, FL, IL, IN, KY, MD,SC, and TX.