Triathlon Revision Knee Catalog #5545-A-301 and Catalog #5546-A-601 The knee system devices are sterile, single-use devices intended for us in revision total knee arthroplasty to alleviate pain and restore function.

FDA Recall Z-1462-2018 · Posted None

Recall StatusTerminated
Product CodeMBH
Date Initiated2018-01-30
Date Terminated2020-05-07
Product Quantity77 total products

Reason for Recall

Customers may receive the incorrect size implant from what is labeled on the box.

Firm Action

On January 31, 2018 branches/agencies received notification of the recall via e-mail and were asked to quarantine affected devices. Notification Letters and Product Accountability Forms were sent to branches, agencies & hospital risk management via UPS (with return receipt) on January 31, 2018 with an expected arrival date of February 02, 2018. Customers were asked to complete a Notification Business Reply Form and return via e-mail or fax. Affected product is to be returned to the manufacturer. Customers with questions may contact (201)831-6693.

Root Cause

Under Investigation by firm

Distribution Pattern

USA (nationwide) Distribution to the states of : AL, AZ, CA, FL, GA, KS, MA, MD, MI, NC, NJ, NY, OH, PA, SC, TN, TX, VA, and WA