ZEISS Axio Observer 5 microscope, Product material no. 431006-9901-000
FDA Recall Z-1465-2022 · Posted None
Advertisement
Recalling FirmZeiss, Carl Inc
Recall StatusTerminated
Product CodeMTX
Date Initiated2022-06-24
Date Terminated2025-02-07
Product Quantity1 US; 2 OUS
Reason for Recall
An incorrect Unique Device Identification (UDI) label was installed on some ZEISS Axio Observer 5 microscopes.
Firm Action
On June 24, 2022, the firm notified their US customer of the recall via email.
The customer was informed that a Zeiss service technician will contact them to arrange a date for performing a corrective action to eliminate the risk. The UDI label will be replaced.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the state of NY.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.