Open, Classified

Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 Medline DIALYSIS CHANGE KIT SKU EBSI1453A Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741

FDA Recall Z-1465-2026 · Posted 2026-02-24

Recall StatusOpen, Classified
Product CodeLRO
Date Initiated2026-01-08
Product Quantity14,525

Reason for Recall

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Firm Action

On January 8, 2026, the firm notified affected customers of the product issue through email and first class mail. Customers were instructed to request stickers to over-label affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No kits are to returned.

Root Cause

Process control

Distribution Pattern

US Nationwide distribution.